Lange split laboratory syringe pump TJ-1A/L0107-1A
Price:Discuss personally

Applicable places:
Biological laboratory use
Product Features:
TJP-3A/W0109-1B Micro-syringe pumps are mainly used in laboratories, and the controller and actuator are divided into structures. It provides a programming working mode, which can perform single-process or multi-process combinations such as multi-stage equal speed, multi-stage mixing curve, trigger allocation, periodic allocation, etc. TJP-3A/W0109-1B Micro-injection pumps are available TJ-3A All functions in TJ-3A Programming mode is added on the basis of the syringe pump.
Functional Features
Custom syringe: can store four user-defined syringe inner diameters
Can you select the syringe from the list or directly enter the syringe inner diameter to determine the syringe?
Linear thrust output: full range >90N
Signal output function: two channels OC Door switch signal output, indicating the start of the pump /Stop and direction
Flow correction function: obtain more accurate liquid volume through calibration operation
Syringe protection function: By adjusting the position of the limit block, the syringe pump can be protected from damage
Block into (retreat )Function: Filling or drawing liquid at full speed
The 485 communication bus can be connected to the upper computer and control it through the background.
External control input interface: 1. External startup /Stop input 2. External event trigger input in programming mode
Control output interface: 1. Two channels OC Door switch signal output, indicating the start of the pump /Stop and Direction 2. State output, only suitable for programming mode
TJP-3A Programming functions of micro-syringe pumps:
In programming mode, you can try to complete a single specific function in a single step, and you can use multiple steps to complete more complex process control functions.
In programming mode, you can choose a variety of solutions such as average speed, growth speed, deceleration, cycle allocation and trigger allocation.
Technical Indicators
Working mode: injection, draw, injection first, then draw, first draw, then injection, continuous
Number of execution units: 1
Maximum itinerary: 90 mm
Travel resolution: 0.165 μm
Linear speed range: 7.94 μm/min-79.4mm/min(flow =Linear speed ×Internal cross-sectional area of the syringe )
Linear speed adjustment resolution: 7.94 μm/min
Stroke control accuracy: Error ≤±0.5%(journey ≥When the maximum itinerary is 30% )
Rated linear thrust: >90N
Syringe selection: built-in main manufacturers and main models of syringes for selection
Syringe customization: can store four user-defined syringe inner meridian values
Flow calibration: Get more accurate liquid volume through the calibration procedure
Operation parameter settings: liquid allocation, injection time, extraction time, interval time, allocation times, etc. Each channel can be set separately, so that all channels can use the parameters of the specified channel.
Start parameter settings: each execution unit starts separately, each execution unit starts delayed, and each execution unit starts synchronously
Display parameter selection: liquid volume, flow rate or linear speed
Power down memory: After powering on again, you can choose whether to continue working according to the state before power down.
fast forward (retreat )Function: Inject or draw liquid at full speed
Status signal output: 2 channels OC Gate signal output, used to indicate activation /Stop and direction status
Control signal input: 2 start and stop control inputs, 1 fallen edge trigger signal control start and stop, 1 TTL Level signal control start and stop
Communication interface: RS485
Controller size: 170 ×108×65(mm)
Controller weight: 0.5 kg
Exemplary dimensions of the execution unit: 245 ×100×95(mm)
Execution unit weight: 1.3 kg
Use power supply: AC 90V-260V/10W
Working ambient temperature: 0 ℃-40℃
Relative humidity of the working environment: <80%
Other parameters
Controller model (Product number ) | Supporting execution unit model (Product number ) | Sampler specifications (ml) | Syringe inner diameter (mm) | Reference flow range (μl/min-ml/min) |
TJP-3A | W0109-1B | 1 | 4.72 | 0.139-1.39 |
2 | 9.00 | 0.505-5.05 | ||
5 | 13.10 | 1.07-10.7 | ||
10 | 16.60 | 1.718-17.18 | ||
20 | 19.00 | 2.251-22.51 | ||
30 | 23.00 | 3.298-32.98 | ||
60 | 29.14 | 5.295-52.95 |