Product/Brand

Multi-Parameter Water Quality Analyzer
Three letters MP521 Type laboratory pH/Conductivity meter
Price:¥4780

WeChat

Brand:SANXIN
Model:MP521
Inventory Status:Spot promotion/please call for consultation
Contact:
Tel:+86 010-51660436
Email:dnyc_sales@163.com
Details Documentation Related

Product Features

1、The most useful pH+Conductivity dual-parameter measuring instrument.

2、Comply with international standards GLP Required, automatic calibration, automatic temperature compensation, data storage, RS232 Intelligent functions such as output, clock display, function settings and self-diagnosis information.

3、There are three different measurement modes: stable display mode; Timing measurement mode ; Continuous measurement mode.

4、Automatic recognition pH Standard buffer solution, there are three series of standard buffer solutions to choose from: European and American series, NIST Series and Chinese series.

5、Adopt advanced conductivity measurement technology, using constants K=1The conductive electrode can meet 0.5 with just a little calibration μS/cm~200mS/cm measurement accuracy requirements.

6、Switchable conductivity, TDS、 Salinity and resistivity. TDS The sum of salinity is calculated using a polynomial to ensure the conversion accuracy of the entire scale.

7、The instrument has a unique product serial number.

8、conform to IP54 Dust-proof and splash-proof grade, the instrument socket is sealed with silicone cap.

 

 Technical parameters

pH

Measurement range

Resolution

Accuracy

Input current

Input impedance

stability

Temperature compensation range

(-1.99~19.99)pH

0.1/0.01pH

Electrical meter: ±0.01pH, Supporting: ±0.02pH

≤1×10-12A

≥3×1012Ω

±0.01pH/3h

(0~100)℃(Automatic or manual)

mV

Measurement range (mV/ORP/EH)

Resolution

Accuracy

-1999mV~0~1999mV

1mV

±0.1%FS

Conductivity

Measurement range 

Resolution

Accuracy

Temperature compensation range

Electrode constant

Reference temperature

Conductivity: (0 ~200)mS/cm, Divided into five ranges: (0.00 ~20.00)μS/cm(20.0~200.0)μS/cm(200~2000)μS/cm

(2.00~20.00)mS/cm(20.0~200.0)mS/cm

Resistivity: (0 ~100)MΩ·cm

TDS:(0~100)g/L

Salinity: (0 ~100)ppt

0.01/0.1/1μS/cm0.01/0.1mS/cm

Electrical meter: ±1.0%FS, Supporting: ±1.50%FS

(0~50)℃(automatic)

0.1//1/10/cm-1

25℃

temperature

Measurement range

Resolution

Accuracy

-10℃~110℃

0.1℃

5~60℃scope: ±0.5℃Remaining range: ±1℃

Other technical parameters

Data storage

Save content

power supply

Communication interface

Size and weight

Quality and safety certification

900Group

Measured value number, measured value, temperature value, ATC or MTC Status, measurement date, measurement time

DC9V/300mA

RS232

160×190×70mm/880g

ISO9001:2000,CE and CMC

Working conditions

Ambient temperature

Ambient humidity

IP grade

5~35℃

≤85%

IP54 Dust-proof and splash-proof

 