Product Features

1、Comply with international standards GLP Required, automatic calibration, automatic temperature compensation, data storage, RS232 Intelligent functions such as output, clock display, function settings and self-diagnosis information.

2、There are three different measurement modes: stable display mode; Timing measurement mode ; Continuous measurement mode.

3、Automatically identify 13 types pH Standard buffer solution, there are four series of standard buffer solutions to choose from: European and American series, NIST Series, Chinese series and custom solutions.

4、Can set high-purity water pH Measuring mode and ammonia-added pure water pH The measurement mode meets the use requirements of the electricity, petrochemical and other industries.

5、have EH Redox potential measurement mode, directly displaying the potential relative to the standard hydrogen electrode ORP value.

6、The dissolved oxygen measurement mode has the functions of automatic temperature compensation, automatic salinity compensation and automatic air pressure compensation.

7、The polarographic dissolved oxygen electrode with a new structure can measure temperature, salinity and dissolved oxygen simultaneously. It is equipped with a dedicated dissolved oxygen electrode calibration sleeve, and the electrode polarization requires only 3 ~5min。

8、The instrument has a unique product serial number.

9、conform to IP54 Dust-proof and splash-proof grade, the instrument socket is sealed with silicone cap.

 

Technical parameters

pH

Measurement range

Resolution

Accuracy

Input current

Input impedance

stability

Temperature compensation range

(-1.999~19.999)pH

0.1/0.01/0.001pH

Electrical meter: ±0.002pH, Supporting: ±0.01pH

≤1×10-12A

≥3×1012Ω

±0.002pH/3h

(0~100)℃(Automatic or manual)

mV

Measurement range (mV/ORP/EH)

Resolution

Accuracy

-1999.9mV~0~1999.9mV

0.1mV

±0.03%FS

Dissolved oxygen

Measurement range

Resolution

Accuracy

Response time

Residual current

Temperature compensation range

Salinity compensation range

Air pressure compensation range

Automatic calibration

Electrode type

(0~20.00)mg/L(ppm)(0~200.0)%

0.1/0.01mg/L(ppm)1/0.1%

Electrical meter: ±0.10mg/L, Supporting: ±0.40mg/L

≤30s(25℃,90%response)

≤0.1mg/L

(0~45)℃(automatic)

(0~45)ppt( automatic)

(80~105)kPa( automatic)

Air saturated with water; Water saturated by air

Polarographic

temperature

Measurement range

Resolution

Accuracy

-10℃~110℃

0.1℃

5~60℃scope: ±0.4℃Remaining range: ±0.8℃

Other technical parameters

Data storage

Save content

 power supply

Communication interface

Size and weight

Quality and safety certification

900Group

Measured value number, measured value, temperature value, ATC or MTC Status, measurement date, measurement time

DC9V/300mA

RS232

160×190×70mm/880g

ISO9001:2008,CE and CMC

Working conditions

Ambient temperature

Ambient humidity

IP grade

5~35℃

≤85%

IP54 Dust-proof and splash-proof